The Fund supports networks of state health policy decision makers to help identify, inspire, and inform policy leaders.
The Milbank Memorial Fund supports two state leadership programs for legislative and executive branch state government officials committed to improving population health.
The Fund identifies and shares policy ideas and analysis to advance state health leadership, strong primary care, and sustainable health care costs.
Keep up with news and updates from the Milbank Memorial Fund. And read the latest blogs from our thought leaders, including Fund President Christopher F. Koller.
The Fund publishes The Milbank Quarterly, as well as reports, issues briefs, and case studies on topics important to health policy leaders.
The Milbank Memorial Fund is is a foundation that works to improve population health and health equity.
September 2017 (Volume 95)
Quarterly Article
Jay G. Ronquillo
Diana M. Zuckerman
Dec 4, 2024
Nov 5, 2024
Oct 30, 2024
Back to The Milbank Quarterly
Policy Points:
Context: Medical software has become an increasingly critical component of health care, yet the regulatory landscape for digital health is inconsistent and controversial. To understand which policies might best protect patients, we examined the impact of the US Food and Drug Administration’s (FDA’s) regulatory safeguards on software-related technologies in recent years and the implications for newly passed legislative changes in regulatory policy.
Methods: Using FDA databases, we identified all medical devices that were recalled from 2011 through 2015 primarily because of software defects. We counted all software-related recalls for each FDA risk category and evaluated each high-risk and moderate-risk recall of electronic medical records to determine the manufacturer, device classification, submission type, number of units, and product details.
Findings: A total of 627 software devices (1.4 million units) were subject to recalls, with 12 of these devices (190,596 units) subject to the highest-risk recalls. Eleven of the devices recalled as high risk had entered the market through the FDA review process that does not require evidence of safety or effectiveness, and one device was completely exempt from regulatory review. The largest high-risk recall categories were anesthesiology and general hospital, with one each in cardiovascular and neurology. Five electronic medical record systems (9,347 units) were recalled for software defects classified as posing a moderate risk to patient safety.
Conclusions: Software problems in medical devices are not rare and have the potential to negatively influence medical care. Premarket regulation has not captured all the software issues that could harm patients, evidenced by the potentially large number of patients exposed to software products later subject to high-risk and moderate-risk recalls. Provisions of the 21st Century Cures Act that became law in late 2016 will reduce safeguards further. Absent stronger regulations and implementation to create robust risk assessment and adverse event reporting, physicians and their patients are likely to be at risk from medical errors caused by software-related problems in medical devices.
Keywords: medical device recalls, FDA, patient safety, software, electronic medical records, health information technology, health IT.
Read on Wiley Online Library
Volume 95, Issue 32 (pages 535–553) DOI: 10.1111/1468-0009.12278 Published in 2017